3rd Generation | 4th Generation | 5th Generation | 6th Generation | 7th Generation | |
Geometry | Rotate-Rotate | Rotate-Stationary | Stationary-Stationary - Imatron electron beam scanner | Rotate-Rotate Helical CT | Multiple detector array CT - rotate rotate geometry |
X-ray Tube | Rotating Tube 360 degree and then reverse | Rotating tube 360 degree and reverse | Large arc of tungsten encircles patient and lies opposite detector ring | Slip ring technology allow continuous movement around gantry | Slip rings allow continuous movement around gantry |
Beam | Wide Beam to form an arc over patient | Wide Beam to form an arc over patient | X-rays produced from focal track as high energy electron beam strikes tungsten - electron beam produced in a cone like structure is electronically steered around the patient | X-ray beam moves in helical pattern | Wide beam |
Detectors | Rotating array of detectors rotating with rotating x-ray tube in Gantry (800) | Stationary ring of detectors (4800+) | Beam on High voltage target ring that covers a 210 degree arc | 512 slice (new 640 slice MDCT), Detector array closely abutted | |
Detector type | Xenon gas (25 atm) or Solid State | Solid State | Solid State | Solid State | Solid State |
Fan Beam Geometry | Source fan - x-ray tube apex | Detector Fan - detector apex | |||
Reference beam measured by | detectors near the edge of the detector array | measured by same detector used for transmission measurement | |||
Slice thickness determined by | Collimator | Collimator | Collimator | Collimator | Slice thickness limited by detector size, increased by binning different # of detector subunits together and moving collimator to the edge of all 4 slices |
Collimator < detector size | Yes | Yes | Yes | Yes | Collimator space 4x > detector array size |
Acquisition | Acquire data while table is moving. Interpolation of helical data set. Interleaving reconstruction (<2:1 clinical limitation). | ||||
Movement artefact | More | more | None | Single breath hold for entire abdomen <30s | Reduce patient movement - acquisition in single breath hold |
Helical Pitch | Collimator pitch = table movement / collimator width | Detector pitch = table movement / detector width | |||
Other | Ring Artifact from detector failure | No Moving Parts Cine CT Cardiac imaging | Use less contrast. Increase patient throughput | 3D images, CT Angiograms and even Image beating heart | |
Speed | <5s / slice | <2s / slice | <50 us | 30s / abdomen series | <0.4 s |
Filter | Bow-tie | Bow-tie | |||
Slice sensitivity profile | Slightly Broader |