CT GENERATIONS COMPARISON


  3rd Generation 4th Generation 5th Generation 6th Generation 7th Generation
Geometry Rotate-Rotate Rotate-Stationary Stationary-Stationary - Imatron electron beam scanner Rotate-Rotate Helical CT Multiple detector array CT - rotate rotate geometry 
X-ray Tube  Rotating Tube 360 degree and then reverse  Rotating tube 360 degree and reverse Large arc of tungsten encircles patient and lies opposite detector ring Slip ring technology allow continuous movement around gantry Slip rings allow continuous movement around gantry
Beam  Wide Beam to form an arc over patient  Wide Beam to form an arc over patient  X-rays produced from focal track as high energy electron beam strikes tungsten - electron beam produced in a cone like structure is electronically steered around the patient X-ray beam moves in helical pattern Wide beam 
Detectors  Rotating array of detectors rotating with rotating x-ray tube in Gantry (800) Stationary ring of detectors (4800+) Beam on High voltage target ring that covers a 210 degree arc   512 slice (new 640 slice MDCT), Detector array closely abutted
Detector type Xenon gas (25 atm) or Solid State Solid State Solid State Solid State Solid State
Fan Beam Geometry Source fan - x-ray tube apex Detector Fan - detector apex      
Reference beam measured by detectors near the edge of the detector array measured by same detector used for transmission measurement      
Slice thickness determined by Collimator Collimator Collimator Collimator Slice thickness limited by detector size, increased by binning different # of detector subunits together and moving collimator to the edge of all 4 slices
Collimator < detector size Yes Yes Yes Yes Collimator space 4x > detector array size
Acquisition       Acquire data while table is moving. Interpolation of helical data set. Interleaving reconstruction (<2:1 clinical limitation).   
Movement artefact More more None Single breath hold for entire abdomen <30s Reduce patient movement - acquisition in single breath hold
Helical Pitch        Collimator pitch = table movement / collimator width Detector pitch = table movement / detector width
Other Ring Artifact from detector failure   No Moving Parts
Cine CT
Cardiac imaging
Use less contrast.
Increase patient throughput
3D images, CT Angiograms and even Image beating heart
Speed <5s / slice <2s / slice <50 us 30s / abdomen series <0.4 s
Filter Bow-tie Bow-tie      
Slice sensitivity profile       Slightly Broader   